NIR reflectance measurements (not for Cary 4000)
Note
Parameters activated by a radio button that are not specifically mentioned in the
following procedure should be set to ’Off (e.g. the Signal-to-noise mode radio
button). These parameters will not affect the procedure.
1. Click the Windows Start button, then Programs, then Cary WinUV, and then Align.
2. Click the Setup button.
3. Set the following parameters:
Cary tab
X mode: Mode Nanometers
X mode: Start 2000
X mode: Stop 300
Y mode: Mode % R
Y mode: Y min 0.00
Y mode: Y max 100.00
Show status display ON
Options tab
SBW/Energy: Beam
mode
Double
SBW/Energy: Slit
height
Reduced
Source/Detector:
Lamps
UV-Vis
Source: Source
changeover (nm)
350.0
Source: Detector
changeover (nm)
800.0
Independent tab
Independent control ON
Measurement mode Auto
UV-Vis controls
Ave time (s) 0.1